The founders were confident: their software was Class IIa, certified, done. The intended purpose said the product supports clinicians by displaying risk information. Then we opened the app. Patients appeared in a list, sorted by urgency, with the most critical highlighted in red at the top.

That list is a clinical decision. And under the MDR, the class of a software product follows the decisions it influences.

Rule 11 in plain words

Software intended to provide information used to take decisions for diagnostic or therapeutic purposes is at least Class IIa. If those decisions may cause a serious deterioration of health or a surgical intervention, it is Class IIb. If they may cause death or an irreversible deterioration, it is Class III. Software that monitors physiological processes is Class IIa, and Class IIb where variations in vital parameters could put the patient in immediate danger. Everything else is Class I.

The word that matters is intended. Under the MDR, the intended purpose comes from the label, the instructions for use and the promotional and sales material as well as the technical documentation, so the website, the pitch deck for hospitals and the user interface all count.

The interface makes the claim

Displaying a value is one thing. Ranking patients, colour-coding urgency, suggesting a next step or flagging who should be seen first is another. These features are often added late, because customers ask for them, and nobody goes back to the classification rationale.

We compare three documents side by side: the intended purpose in the technical file, the sales material and the product itself. When they describe three different products, the notified body will eventually notice, usually at the worst possible moment.

What a higher class costs

Moving up a class means a deeper clinical evaluation, often new clinical data, a broader scope on the notified body certificate and more scrutiny of software development under IEC 62304 and risk management under ISO 14971. Notified body capacity is still tight, so the timeline is commonly well over a year. For a company that planned to launch in new markets next quarter, that is a different business plan.

On top of the MDR, the EU AI Act classifies AI-based medical devices that need a notified body as high-risk systems. From August 2028, after the delay agreed in the 2026 Digital Omnibus, they have to meet additional requirements on data governance, transparency, human oversight and logging. Few roadmaps we read have budgeted for it.

Questions we ask management

A class upgrade rarely kills a company. Finding out about it after closing is expensive.

Frequently asked questions

What is MDR Rule 11?

Rule 11 of the EU Medical Device Regulation classifies software. Software that informs diagnostic or therapeutic decisions is at least Class IIa, Class IIb if wrong decisions could cause serious deterioration of health or surgery, and Class III if they could cause death or irreversible harm.

Can a software update change the MDR class?

Yes. A new feature, such as ranking patients or recommending treatment, can change the decisions the software informs and with them the class. The classification rationale should be reviewed with every such change.

Our guide to healthcare due diligence covers all four areas of a review.